(A) A licensee shall maintain a record of the calibrations of brachytherapy sources required by rule 3701:1-58-48 of the Administrative Code for three years after the last use of the source.
(B) The record must include:
(1) The date of the calibration;
(2) The manufacturer's name, model number, and serial number for the source and the instruments used to calibrate the source;
(3) The source output or activity;
(4) The source positioning accuracy within the applicators; and
(5) The name of the individual, the source manufacturer, or the calibration laboratory that performed the calibration.
R.C. 119.032 review dates: 07/12/2010 and 07/12/2015
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04