3701-84-33 Safety standards.

(A) The provider of a cardiac catheterization service shall establish and maintain safety guidelines, and practices and policies in accordance with applicable United States nuclear regulatory commission regulations, applicable provisions of Chapter 3748. of the Revised Code, and the rules adopted pursuant to that chapter to assure a safe environment for patients, visitors, and personnel.

(B) The provider of a cardiac catheterization service shall establish, maintain and follow electrical safety policies that, at a minimum, include:

(1) A safe primary electrical wiring system;

(2) Electrical isolation of all equipment attached to a patient;

(3) Use of an equipotential hardwired grounding system for all equipment; and

(4) Periodic inspection of the electrical system and measurement of interequipment current leakage.

(C) The provider of a cardiac catheterization service shall periodically survey all of the equipment utilized by the service and perform preventive maintenance on a schedule that, at a minimum, conforms to manufacturers' recommendations. Results of surveillance and preventive maintenance activities shall be internally documented.