Mobile Intensive Care Units licensed at the pediatric level must carry adult, pediatric, & neonatal equipment and supplies listed in rules 4766-4-21, 4766-4-22, and 4766-4-23 of the Administrative Code.
(A) Isolation equipment
(1) Four packaged kits or
(a) Four isolation goggles & masks or mask/shield combination;
(b) Four isolation gowns;
(c) Four isolation gloves;
(2) Four isolation high particulate (HEPA)-assorted sizes;
(3) Four containers (bags) for infectious medical waste;
(4) Sharps container;
(5) Disinfectant/germicidal;
(6) Waterless hand cleaner.
(B) Airway equipment
(1) Complete oropharyngeal airway kit-adult, pediatric, and neonatal;
(2) Complete nasopharyngeal airway kit-adult, pediatric, and neonatal;
(3) Complete intubation kit-adult, pediatric, and neonatal:
(a) Extra batteries and bulbs;
(b) Syringes, assorted sizes;
(c) Adult stylet;
(d) Pediatric stylet;
(e) Neonatal stylet;
(f) Adult magill forceps;
(g) Pediatric magill forceps;
(h) Booted hemostat;
(i) Adult endotracheal tube set;
(j) Pediatric endotracheal tube set;
(k) Water soluble lubricant;
(l) Laryngoscope handle;
(m) Laryngoscope blades, curved and straight, sizes 0-1-2-3;
(n) End-tidal CO2 detector;
(4) Invasive airway kit, per protocol, (i.e. cricothyrotomy set, pertrach set);
(5) Two ventilator circuits;
(6) Advanced airway-(i.e. laryngeal mask airway) pediatric and adult.
(C) Definitive equipment
(1) Approved medications, per protocol;
(2) ECG monitor/defibrillator and appropriate pads, adult/pediatric:
(a) One set pediatric paddles;
(3) Hemodynamic monitoring equipment;
(4) Pulse oximeter;
(5) Doppler & gel;
(6) Two I.V. pumps;
(7) Ventilator with minimum of PEEP;
(8) Generator and 110 power source;
(9) Chest tube draining system.
(D) Bleeding/burns equipment
(1) Ten adhesive dressing strips;
(2) Twenty gauze pads, assorted sizes;
(3) Four surgi pads/sterile sponge pads;
(4) Four assorted standard kling/kerlix;
(5) Two universal trauma dressings, size 10” by 36”;
(6) Two sterile burn sheets or two burn packs;
(7) Four sterile water and/or .9% n.s. for irrigation.
(E) Suction equipment
(1) Wall mounted suction unit;
(2) Portable suction unit, battery operated;
(3) Two hard tip suction (Yankauer);
(4) Soft tip suction catheters set, adult & pediatric sizes;
(5) Two suction tubing;
(6) Suction bags (package) or equivalent;
(7) Two pairs sterile gloves.
(F) Oxygen equipment
(1) Main oxygen (M tank or greater);
(2) Wall mounted oxygen gauge 0-15 L/min. minimum;
(3) Compress air as appropriate;
(4) Two portable oxygen unit – minimum “D” tanks;
(5) Portable variable flow regulator 0-15 L/min. minimum;
(6) Bag-valve-mask with reservoir 100% oxygen flow:
(a) Two adult;
(b) Two pediatric;
(c) Two neonatal;
(7) Transparent oxygen masks, simple and non rebreather:
(a) Two adult;
(b) Two pediatric;
(c) Two neonatal or head hoods;
(8) Nasal cannulas:
(a) Two adult;
(b) Two pediatric;
(c) Two neonatal;
(9) Oxygen connective tubing and appropriate adapters;
(10) Two oxygen humidifier/nebulizer and appropriate connecting tubing.
(G) Adjunct equipment
(1) Scissors;
(2) Stethoscope-adult, pediatric, and infant;
(3) B/P cuffs:
(a) Neonatal, sizes 1,2,3,4;
(b) Infant;
(c) Pediatric;
(d) Adult;
(e) Large adult;
(4) Thermometer;
(5) Penlight;
(6) Flashlight;
(7) Two tongue blades;
(8) Two bite sticks;
(9) Patient cot with three straps/isolette with two straps;
(10) Two eye shields or eye protector pads;
(11) Four assorted tape;
(12) Exam gloves, one box each of small-medium-large or one box of uni-size;
(13) Two emesis basins or equivalent;
(14) Obstetrical kit;
(15) Urinal or equivalent;
(16) Bed pan or equivalent;
(17) Two ABC fire extinguisher, 5 lb. minimum, front and back;
(18) Nasogastric tubes, adult & pediatric sizes;
(19) Four patient restraints;
(20) Two each of hot & cold packs;
(21) Two pillows & cases;
(22) Six sheets;
(23) Two blankets.
(H) Intravenous equipment
(1) Ten alcohol or betadine preps;
(2) Two short arm boards;
(3) Four IV administration sets:
(a) Four IV infusion pump tubing;
(4) IV catheters and butterfly needles, assorted sizes;
(5) Two intraosseous needles;
(6) Needles, assorted sizes;
(7) Two IV solutions, per protocol;
(8) Associated adjunct equipment:
(a) Two invasive line set-up;
(b) Two pressure bags.
Effective: 04/18/2005
R.C. 119.032 review dates: 04/18/2010
Promulgated Under: 119.03
Statutory Authority: 4766.03
Rule Amplifies: 4766.04, 4766.07