Skip to main content
Back To Top Top Back To Top
This website publishes administrative rules on their effective dates, as designated by the adopting state agencies, colleges, and universities.

Rule 3701:1-58-69 | Five-year inspection for teletherapy and gamma stereotactic radiosurgery units.

 

(A) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during source replacement, to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed five years for each teletherapy unit and shall not exceed seven years for each gamma stereotactic radiosurgery unit.

(B) This inspection and servicing may only be performed by persons specifically licensed to do so by the director, the United States nuclear regulatory commission, or an agreement state.

(C) A licensee shall keep a record of the inspection and servicing in accordance with rule 3701:1-58-100 of the Administrative Code.

Last updated November 18, 2022 at 8:39 AM

Supplemental Information

Authorized By: 3748.02, 3748.04
Amplifies: 3748.04
Five Year Review Date: 8/15/2026
Prior Effective Dates: 8/15/2005, 10/4/2010, 8/10/2015