Skip to main content
Back To Top Top Back To Top
This website publishes administrative rules on their effective dates, as designated by the adopting state agencies, colleges, and universities.

Rule 3701:1-58-96 | Records of periodic spot-checks for remote afterloader units.

 

(A) A licensee shall retain a record of each spot-check for remote afterloader units required by rule 3701:1-58-65 of the Administrative Code for three years.

(B) The record must include, as applicable:

(1) The date of the spot-check;

(2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source;

(3) An assessment of timer accuracy;

(4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and

(5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.

(C) A licensee shall retain a copy of the procedures required by paragraph (B) of rule 3701:1-58-65 of the Administrative Code until the licensee no longer possesses the remote afterloader unit.

Last updated November 18, 2022 at 8:42 AM

Supplemental Information

Authorized By: 3748.02, 3748.04
Amplifies: 3748.04
Five Year Review Date: 4/7/2027