Rule 3701-84-78 | Safety standards - pediatric cardiac catheterization service.

 

(A) To assure a safe environment for patients, visitors, and personnel, each pediatric cardiac catheterization service will establish and maintain safety guidelines, practices and policies in accordance with applicable United States nuclear regulatory commission regulations, applicable provisions of Chapter 3748. of the Revised Code, and the rules adopted pursuant to that chapter.

(B) Each pediatric cardiac catheterization service will establish, maintain, and follow electrical safety policies which include:

(1) A safe primary electrical wiring system;

(2) Electrical isolation of all equipment attached to a patient;

(3) Use of an equipotential hardwired grounding system for all equipment; and

(4) Periodic inspection of the electrical system and measurement of interequipment current leakage.

(C) Each pediatric cardiac catheterization service will periodically survey all of the equipment utilized by the service, perform preventive maintenance on a schedule that conforms to manufacturers' recommendations, and document all preventive maintenance activities.